Allure Quadra Adverse Events: FDA MAUDE Data
ABBOTT (ST. JUDE MEDICAL) · CRT-Defibrillator (CRT-D) · First cleared May 2004 · 2 device variants grouped
Total Reports
347
Reports (30d)
0
Active Signals
1
Open Recalls
2
Embed this chart in your article
Free to use with the source line. The chart updates weekly.
Detection
Active safety signals
Regulatory
FDA recalls for this device
Aug 16, 2023 · Open, Classified
Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.
Mar 10, 2022 · Open, Classified
Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.
Evidence
Related research
Profile of St. Jude Medical's Allure Quadra quadripolar pacemaker system for cardiac resynchronization therapy.
Corbisiero R, et al. · Expert review of medical devices · 2014
Case of Frozen Leads: An Old but Persistent Challenge
Takamasa Ishikawa, et al. · Pacing and Clinical Electrophysiology · 2025
Cardiac Resynchronization Therapy Delivered Via a Multipolar Left Ventricular Lead is Associated with Reduced Mortality and Elimination of Phrenic Nerve Stimulation: Long-Term Follow-Up from a Multicenter Registry.
Behar JM, et al. · Journal of cardiovascular electrophysiology · 2015
Visualization of the SyncAV® Algorithm for CRT Optimization by Non-invasive Imaging of Cardiac Electrophysiology: NICE-CRT Trial
Philipp Spitaler, et al. · Journal of Clinical Medicine · 2023
Corynebacterium striatum Cardiac Device-Related Infective Endocarditis: The First Case Report in a Patient With a Cardiac Resynchronization Therapy Defibrillator Device and Review of the Literature
Luísa Serpa Pinto, et al. · Journal of Medical Cases · 2020
Citations indexed from PubMed.
Keep exploring
Related devices in CRM
CapSureFix Novus
MEDTRONIC, INC. · 67,431 reports
Emblem S-ICD
BOSTON SCIENTIFIC CORPORATION · 39,451 reports
Sprint Quattro Secure S MRI SureScan
MEDTRONIC, INC. · 39,199 reports
Durata
ABBOTT (ST. JUDE MEDICAL) · 20,793 reports
Quadra Assura MP
ABBOTT (ST. JUDE MEDICAL) · 10,759 reports
LINQ II
MEDTRONIC, INC. · 10,515 reports