Endotak Reliance S Adverse Events: FDA MAUDE Data
BOSTON SCIENTIFIC CORPORATION · Permanent Defibrillator Electrode (Lead) · First cleared Sep 2016 · 4 device variants grouped
Total Reports
1,106
Reports (30d)
8
Active Signals
0
Open Recalls
4
Embed this chart in your article
Free to use with the source line. The chart updates weekly.
What the reports describe · Claripulse analysis
Most commonly reported problems
Right Ventricular Lead Explant for Infection
explanted right, infection additional, effects, revision infection, lead revision
5
reports
Gradual RV Shock Impedance Rise Under Monitoring
impedance measurements, measurements, effects, shock impedance, service
3
reports
RV Lead Dislodgement Requiring Explant
effects, reported right, dislodged, additional adverse, surgically
2
reports
Gradually Rising RV Shock Impedance
defibrillator crt, cardiac resynchronization, crt, therapy defibrillator, resynchronization therapy
1
reports
Implantable Lead Removed for Infection
implantable lead, effects, revision infection, reported implantable, lead revision
1
reports
Right Ventricular Lead Infection Revision
explanted right, infection additional, effects, revision infection, additional adverse
1
reports
Problem groupings are generated by Claripulse from report narrative text. They summarize what reporters wrote, not adjudicated root cause.
Regulatory
FDA recalls for this device
Jul 24, 2025 · Open, Classified
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Jul 24, 2025 · Open, Classified
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Sep 30, 2022 · Open, Classified
There is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
Sep 30, 2022 · Open, Classified
There is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
Keep exploring
Related devices in CRM
CapSureFix Novus
MEDTRONIC, INC. · 67,431 reports
Emblem S-ICD
BOSTON SCIENTIFIC CORPORATION · 39,451 reports
Sprint Quattro Secure S MRI SureScan
MEDTRONIC, INC. · 39,199 reports
Durata
ABBOTT (ST. JUDE MEDICAL) · 20,793 reports
Quadra Assura MP
ABBOTT (ST. JUDE MEDICAL) · 10,759 reports
LINQ II
MEDTRONIC, INC. · 10,515 reports