Total Reports

248

Reports (30d)

0

Active Signals

1

Open Recalls

4

Monthly Report Volume
Event Type Breakdown

What the reports describe · Claripulse analysis

Most commonly reported problems

1

Device Disconnected from Remote Monitoring

observed, reprogramming, event patient, stable, technical

1

reports

Problem groupings are generated by Claripulse from report narrative text. They summarize what reporters wrote, not adjudicated root cause.

Detection

Active safety signals

Disproportionate reportinghigh

Regulatory

FDA recalls for this device

Class II

Aug 16, 2023 · Open, Classified

Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

Class II

Mar 10, 2022 · Open, Classified

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Class II

Mar 10, 2022 · Open, Classified

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Class II

Mar 10, 2022 · Open, Classified

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Evidence

Related research

Citations indexed from PubMed.

Keep exploring

Related devices in CRM

CAPSUREFIX Novus

MEDTRONIC, INC. · 43,528 reports

→

EMBLEM S-ICD

BOSTON SCIENTIFIC CORPORATION · 36,855 reports

→

Sprint Quattro Secure S MRI SureScan

MEDTRONIC, INC. · 36,335 reports

→

CapSureFix Novus MRI SureScan

MEDTRONIC, INC. · 27,017 reports

→

Durata

ABBOTT (ST. JUDE MEDICAL) · 19,619 reports

→

Quadra Assura MP

ABBOTT (ST. JUDE MEDICAL) · 10,450 reports

→

Want automated safety signal detection?

Claripulse monitors these devices weekly and detects anomalies using 6 statistical methods — spike, trend, changepoint, peer-outlier, and more. 1 active signal detected right now.

Get Early Access

Understanding this data

What FDA MAUDE data does and does not tell you

How these reports are filed, why raw counts mislead, and the limits to know before drawing conclusions.

→

How a safety signal is detected

What the "Active Signals" count above means and the methods that separate a real signal from noise.

→
Disclaimer: This system analyzes publicly available FDA MAUDE data. It does NOT provide medical advice. MAUDE data is de-identified. Do not attempt re-identification. Report counts may differ from the FDA's MAUDE search due to processing cadence and deduplication. Data is updated weekly.