Micra AV2 Adverse Events: FDA MAUDE Data
MEDTRONIC, INC. · Leadless Pacemaker · First cleared Apr 2023
Total Reports
3,296
Reports (30d)
105
Active Signals
0
Open Recalls
0
Embed this chart in your article
Free to use with the source line. The chart updates weekly.
What the reports describe · Claripulse analysis
Most commonly reported problems
Intraoperative Failure to Capture with High Thresholds
pulse, implantable pulse, pulse generator, generator ipg, ipg
159
reports
Recapture-Induced Perforation with Tamponade
pulse, implantable pulse, pulse generator, generator ipg, ipg
65
reports
Early Post-Implant Intermittent Capture Failure
pulse, implantable pulse, pulse generator, generator ipg, ipg
36
reports
Premature Battery Depletion with High Thresholds
pulse, implantable pulse, pulse generator, leadless implantable, ipg
9
reports
Post-Implant Migration into Atrium
deployment, surgery, high capture, cardiac perforation, completed
9
reports
Premature Tether-Mode Separation After Fixation
left, tether mode, release, tether, prematurely
1
reports
Problem groupings are generated by Claripulse from report narrative text. They summarize what reporters wrote, not adjudicated root cause.
Evidence
Related research
Advancements in Leadless Pacemakers: What the Second-generation Micra AV2 Brings to Cardiac Care
Christophe Garweg, et al. · Heart International · 2024
Sudden Micra VR Battery Failure With Telemetry Loss and Fluoroscopic Battery Deformation
Yuya Nakamura, et al. · Pacing and Clinical Electrophysiology · 2025
Cardiac transplant-associated complete AV block treated with a leadless pacemaker: a case report
Débora Sá, et al. · European Heart Journal - Case Reports · 2025
Combined Lead Extraction and Leadless VDD Pacemaker Implantation in d-TGA Patient After Senning Procedure
Hamza Hamayel, et al. · JACC Case Reports · 2026
Comparative Analysis of Extraction Complications in Transvenous Versus Leadless Pacemakers
Abdullah Sahyouni, et al. · Cureus · 2026
Citations indexed from PubMed.
Keep exploring
Related devices in CRM
CapSureFix Novus
MEDTRONIC, INC. · 67,431 reports
Emblem S-ICD
BOSTON SCIENTIFIC CORPORATION · 39,451 reports
Sprint Quattro Secure S MRI SureScan
MEDTRONIC, INC. · 39,199 reports
Durata
ABBOTT (ST. JUDE MEDICAL) · 20,793 reports
Quadra Assura MP
ABBOTT (ST. JUDE MEDICAL) · 10,759 reports
LINQ II
MEDTRONIC, INC. · 10,515 reports
Want automated safety signal detection?
Claripulse monitors these devices weekly and detects anomalies using 6 statistical methods: spike, trend, changepoint, peer-outlier, and more.
Get Early AccessUnderstanding this data