Valitude X4 CRT-P Adverse Events: FDA MAUDE Data

BOSTON SCIENTIFIC CORPORATION · CRT-Pacemaker (CRT-P) · First cleared Oct 2014

Total Reports

2,270

Reports (30d)

71

Active Signals

0

Open Recalls

17

Monthly Report Volume
Event Type Breakdown
Embed this chart in your article

Free to use with the source line. The chart updates weekly.

What the reports describe · Claripulse analysis

Most commonly reported problems

1

High-Impedance Battery Failure Requiring Replacement

ventricular, lead, replaced, services, crt recorded

48

reports

2

Low-Voltage Alert with High Battery Impedance

recommended device, indicating voltage, reported cardiac, code indicating, crt recorded

16

reports

3

Low-Voltage Alert Pending Battery Analysis

capacity request, indicating voltage, code indicating, device recorded, reported device

12

reports

4

Infection-Driven System Revision and Explant

explanted cardiac, crt revision, infection additional, revision infection, crt explanted

5

reports

5

Low-Voltage Alert with Continued Monitoring

capacity request, indicating voltage, service time, code indicating, crt recorded

5

reports

6

Low-Voltage Alert with Battery Impedance Concern

crt remains, states voltage, states, code states, crt recorded

2

reports

Problem groupings are generated by Claripulse from report narrative text. They summarize what reporters wrote, not adjudicated root cause.

Regulatory

FDA recalls for this device

Class I

Mar 19, 2026 · Open, Classified

Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia

Class I

Aug 20, 2025 · Open, Classified

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Class I

Aug 20, 2025 · Open, Classified

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Class I

Aug 20, 2025 · Open, Classified

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Class I

Aug 20, 2025 · Open, Classified

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Class I

Aug 20, 2025 · Open, Classified

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Want automated safety signal detection?

Claripulse monitors these devices weekly and detects anomalies using 6 statistical methods: spike, trend, changepoint, peer-outlier, and more.

Get Early Access

Understanding this data