VISIONIST CRT-P Adverse Events: FDA MAUDE Data
BOSTON SCIENTIFIC CORPORATION · CRT-Pacemaker (CRT-P) · First cleared Dec 2016 · 2 device variants grouped
Total Reports
629
Reports (30d)
17
Active Signals
0
Open Recalls
17
What the reports describe · Claripulse analysis
Most commonly reported problems
High Impedance Leading to Device Explant
explanted, atrial, right, ventricular, lead
21
reports
High Impedance with Ongoing Monitoring
condition, provided, monitored, implanted patient, time adverse
10
reports
High Impedance with No Patient Effects
7
reports
High Impedance with Device Still in Service
resulted, remain, monitored, implanted patient, time adverse
2
reports
High Impedance with No Adverse Effects
observed, reported pacemaker, atrial, recommended time, pacemaker recorded
2
reports
High Impedance with No Replacement
recommended adverse, today, reported crt, service cardiac, time cardiac
2
reports
Problem groupings are generated by Claripulse from report narrative text. They summarize what reporters wrote, not adjudicated root cause.
Regulatory
FDA recalls for this device
Mar 19, 2026 · Open, Classified
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Evidence
Related research
Citations indexed from PubMed.
Keep exploring
Related devices in CRM
CAPSUREFIX Novus
MEDTRONIC, INC. · 46,344 reports
Sprint Quattro Secure S MRI SureScan
MEDTRONIC, INC. · 39,133 reports
EMBLEM S-ICD
BOSTON SCIENTIFIC CORPORATION · 39,062 reports
CapSureFix Novus MRI SureScan
MEDTRONIC, INC. · 29,710 reports
Durata
ABBOTT (ST. JUDE MEDICAL) · 20,458 reports
Quadra Assura MP
ABBOTT (ST. JUDE MEDICAL) · 10,671 reports
Want automated safety signal detection?
Claripulse monitors these devices weekly and detects anomalies using 6 statistical methods — spike, trend, changepoint, peer-outlier, and more.