VISIONIST CRT-P Adverse Events: FDA MAUDE Data
BOSTON SCIENTIFIC CORPORATION · CRT-Pacemaker (CRT-P) · First cleared Jan 2004
Total Reports
523
Reports (30d)
20
Active Signals
1
Open Recalls
17
What the reports describe · Claripulse analysis
Most commonly reported problems
Lead Failure with Device Explanted
explanted, pacing, right, ventricular, lead
27
reports
Battery High Impedance with Patient Monitoring
remain implanted, device remain, implanted patient, monitored time, patient monitored
7
reports
Battery High Impedance with Service Recommendations
battery condition, capacity reported, recommended adverse, reported request, service time
6
reports
Battery High Impedance with Monitoring
battery condition, capacity data, battery continued
6
reports
Battery High Impedance with Continued Service
recommended adverse, service time, services, service cardiac, battery continued
5
reports
Battery High Impedance with Device Replaced
device explanted, replaced, additional adverse, explanted replaced, explanted
5
reports
Problem groupings are generated by Claripulse from report narrative text. They summarize what reporters wrote, not adjudicated root cause.
Detection
Active safety signals
Regulatory
FDA recalls for this device
Mar 19, 2026 · Open, Classified
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Aug 20, 2025 · Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Keep exploring
Related devices in CRM
CAPSUREFIX Novus
MEDTRONIC, INC. · 44,929 reports
EMBLEM S-ICD
BOSTON SCIENTIFIC CORPORATION · 37,761 reports
Sprint Quattro Secure S MRI SureScan
MEDTRONIC, INC. · 37,430 reports
CapSureFix Novus MRI SureScan
MEDTRONIC, INC. · 28,090 reports
Durata
ABBOTT (ST. JUDE MEDICAL) · 19,969 reports
Quadra Assura MP
ABBOTT (ST. JUDE MEDICAL) · 10,531 reports
Want automated safety signal detection?
Claripulse monitors these devices weekly and detects anomalies using 6 statistical methods — spike, trend, changepoint, peer-outlier, and more. 1 active signal detected right now.
Get Early AccessUnderstanding this data